Ventlab Infant AirFlow
Resuscitator
AF 3040 and AF 3140
| Bag Volume | 300 ml | ||
| Stroke Volume | < 150 ML (ASTM Standard Hand) | ||
| Deadspace | 7 ml with mask or low dead space adapter | ||
| Exhalation Resistance | < 5 cm H2O @ 5 lpm flow | ||
| Pressure Relief (optional) | 40 cm H2O | ||
| Attainable Delivery Pressure | 40 cm H2O and override | ||
| Ventilatory Frequency | > 90 bpm (NO LOAD) | ||
| Operating Temperature | -18oC to 51oC (0o TO 123oF) | ||
| Storage Environment Limits | -40oC to 60oC (-40o TO 140oF) | ||
| Patient Port Inlet | 15mm/22mm | ||
| Average Device Mass without Mask | 0.20 kg (0.46 lb) | ||
| Average Length (Basic Model) | 22.5 cm (9") | ||
| Body Mass Use Range | < 10 kg (22 lbs) | ||
| Quantity per case | 10 | ||
| TIDAL VOLUME | 150 ml | 150 ml | 150 ml |
| Rate | 15 bpm | 20 bpm | 30 bpm |
| O2 flowrate | 10 lpm | 10 lpm | 10 lpm |
| FDO2 | 99% | 99% | 99% |