Ventlab Infant AirFlow Resuscitator
AF 3040 and AF 3140

Bag Volume  300 ml
Stroke Volume  < 150 ML (ASTM Standard Hand)
Deadspace  7 ml with mask or low dead space adapter
Exhalation Resistance  < 5 cm H2O @ 5 lpm flow
Pressure Relief (optional)  40 cm H2O
Attainable Delivery Pressure  40 cm H2O and override
Ventilatory Frequency  > 90 bpm (NO LOAD)
Operating Temperature  -18oC to 51oC (0o TO 123oF)
Storage Environment Limits  -40oC to 60oC (-40o TO 140oF)
Patient Port Inlet  15mm/22mm
Average Device Mass without Mask  0.20 kg (0.46 lb)
Average Length (Basic Model)  22.5 cm (9")
Body Mass Use Range < 10 kg (22 lbs)
Quantity per case 10
 
TIDAL VOLUME 150 ml 150 ml 150 ml
Rate 15 bpm 20 bpm 30 bpm
O2 flowrate 10 lpm 10 lpm 10 lpm
FDO2 99% 99% 99%